AlarmTH Module — Connectivity (TH ↔ THIB ↔ MEC)

HC11016Thermal printhead data line 2 (P2 to MEC P23) connectivity fault

Agfa CT error code, reported by the TH Module — Connectivity (TH ↔ THIB ↔ MEC).

What HC11016 means

On DryStar 5500 (Type 5364/100) and DryStar 5503 (Type 5364/300) CT systems, status code HC11016 indicates: Thermal printhead data line 2 (P2 to MEC P23) connectivity fault. The code originates in the TH Module — Connectivity (TH ↔ THIB ↔ MEC) section of the TH Module — Connectivity (TH ↔ THIB ↔ MEC). Severity level alarm (Alarm).

Reported action: Check connectors P2 (center of the thermal printhead) and P23 (MEC board) — reconnect if not or badly connected. If the code persists, replace the data cable between P2 and P23; if it still persists after the cable swap, replace the thermal printhead itself..

Raised on the Drystar 5500/5503 when the connectivity of the data cable between P2 on the thermal printhead and P23 on the MEC board is bad, shown on the local display as "Call service".

Full troubleshooting for HC11016

The complete diagnosis, cause tree, and step-by-step remedy for the TH Module — Connectivity (TH ↔ THIB ↔ MEC) subsystem are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with real figures.

Open the Activion16 service manual

Related TH Module — Connectivity (TH ↔ THIB ↔ MEC) codes

Discuss HC11016 with other field engineers in the FSELIB forum

Source: AGFA DryStar service manual — HC error catalogue (FSELIB authored corpus). Reference information re-authored by FSELIB — not a copy of the OEM manual.