HC10903 — Print drum motor M1 not connected or malfunctioning
Agfa CT error code, reported by the Print Drum Module.
What HC10903 means
On DryStar 5500 (Type 5364/100) and DryStar 5503 (Type 5364/300) CT systems, status code HC10903 indicates: Print drum motor M1 not connected or malfunctioning. The code originates in the Print Drum Module — Drive Motor & Mechanics section of the Print Drum Module. Severity level alarm (Alarm).
Reported action: Check the drum drive belts first: verify the flat belt driving the drum is mounted correctly on the pulleys and remount or replace it if not mounted, ruptured, or damaged, then check the toothed drum-motor belt the same way. Next check the connections to motor M1 and reconnect or replace the connector if faulty, then connector XS8, then connector P36 on the MEC board, and verify the wiring from M1 to MEC/P36 with a multimeter. If wiring is good, power off the printer and replace motor M1; if the fault persists, the motor function on the MEC board has failed and the MEC board must be replaced..
Raised on the Drystar 5500/5503 print drum module when the print drum motor M1 is not connected or is malfunctioning. Classed as a Defect, and the printer shows "Call service". Documented in the appendix alongside the drum-positioning motor fault HC10910.
Full troubleshooting for HC10903
The complete diagnosis, cause tree, and step-by-step remedy for the Print Drum Module — Drive Motor & Mechanics subsystem are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with real figures.
Open the Activion16 service manualRelated Print Drum Module — Drive Motor & Mechanics codes
Source: AGFA DryStar service manual — HC error catalogue (FSELIB authored corpus). Reference information re-authored by FSELIB — not a copy of the OEM manual.