HC10050 — Themis CPU board self-test failed
Agfa CT error code, reported by the CPU Boards (Crius/Themis).
What HC10050 means
On DryStar 5500 (Type 5364/100) and DryStar 5503 (Type 5364/300) CT systems, status code HC10050 indicates: Themis CPU board self-test failed. The code originates in the CPU Board (Themis) — Board Test section of the CPU Boards (Crius/Themis). Severity level alarm (Alarm).
Reported action: The source gives a single cause and no field repair: the Themis CPU board (PCB) is not functioning correctly. Power off the printer and replace the Themis board..
Raised on the Drystar 5500/5503 when the self-test of the Themis CPU board fails. Filed in the appendix's CPU-board fault-code table alongside the equivalent Crius board test code, HC10000, and the more specific Themis RAM test, HC10052.
Full troubleshooting for HC10050
The complete diagnosis, cause tree, and step-by-step remedy for the CPU Board (Themis) — Board Test subsystem are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with real figures.
Open the Activion16 service manualRelated CPU Board (Themis) — Board Test codes
Source: AGFA DryStar service manual — HC error catalogue (FSELIB authored corpus). Reference information re-authored by FSELIB — not a copy of the OEM manual.