Saggital Scan Pauses with Gradient Faults 666

Symptom

Sagittal Scan Pauses with Gradient Fault 666 Intermittent gradient power supply faults are occurring during sagittal localizer scan. Axial completes, then scan pauses after sagittal series is started. Restarting the scan in the coronal plane allows normal completion. HV message 666 (HV restart) observed on SCA in PSU during sagittal scan. Image collapses with bright horizontal line in sagittal view. 3-plane localizer protocol stresses gradients. Reducing the FOV will induce failure more easily.

Error message

2240308 A high voltage power supply fault occurred on the Power Supply Unit. Tue Aug 3 12:11:30 2004 Error: 2240308 Host: lx-mr Process: NSP File: HPC_PORT_ID:spi_psu_error Line: 4 Gradient Power Supply Unit Fault. A high voltage power supply fault occurred on the Power Supply Unit. Possible causes: X, Y, or Z axis amplifier has shorted the supply.

2240255 Y Axis Gradient Amplifier Status. The auxiliary voltage is present but the high voltage is off. Tue Aug 3 12:11:39 2004 Error: 2240255 Host: lx-mr Process: NSP File: HPC_PORT_ID:spi_gca_error Line: 5 Y Axis Gradient Amplifier Status. The auxiliary voltage is present but the high voltage is off. Verify the "REM/LOC" switch on front panel of the System Control.

Applies to

Product
Profile 3 Ultrasparc CPU with AN8290Y gradients
Software
Profile Contour 7.67 Applies to Profile1 through 4 and Contour1 through 4. This is the last revision of software for this product line. Requires 7.66 LFC. 7.67 is a patch.
Component
2178378 AN8290Y GCA Gradient Amp Module
Area
MR

Full cause & step-by-step resolution