GE Healthcare · MRSigna22403082224666
Analogic 8280 Gradients Fail to Come Up with HV PS Fault Due to Gradient Power Supply Unit
Symptom
Analogic 8280 Gradients Fail to Come Up with HV PS Fault. HFO and Signa Base (SR20) systems. See Long Description OpenSpeed system running HFO3 M4 software and Signa base (SR20) systems, the gradients fail to come up ready. Have HV PS FLT on the SCA. We measured the HV at the amps and it is pretty close to zero.
Error message
2240308 A high voltage power supply fault occurred on the Power Supply Unit. Tue Aug 3 12:11:30 2004 Error: 2240308 Host: lx-mr Process: NSP File: HPC_PORT_ID:spi_psu_error Line: 4 Gradient Power Supply Unit Fault. A high voltage power supply fault occurred on the Power Supply Unit. Possible causes: X, Y, or Z axis amplifier has shorted the supply. 2224666 Auto Prescan failed, Gradient amplifiers are not ready
Applies to
- Models
- Signa
- Product
- HF OpenSpeed HFO1 HFO2 and HFO3 Only.
- Software
- HFO3
- Component
- 2273028-2 POWER SUPPLY UNIT FOR AN8280G GRAD 2273028-2 POWER SUPPLY UNIT FOR AN8280G GRAD
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- AE title has changed on destination host.
- CWA ID100315. Auto and Manual Prescan Fails.
- What is the Acoustic Noise Level in the Ovation Magnet.