Pular para o conteúdo principal
GE Healthcare · MR22402792240255224026322402712240239

No ID Flt Displayed on the 8290 Gradient SCA Board.

Symptom

No ID Flt Displayed on the 8290 Gradient SCA Board. The SCA display showed a No ID flt message. The gradients were not ready. X and Z were in stand-by (state 5), Y amp had error 13. Several different errors were reported in the system error log. The problem would occur when powering up.

Error message

2240279 Z Axis Gradient Amplifier Fault. IPG received an unknown status 0x13 from the Gradient Subsystem. Verify no errors are reported on the front of the System Control Assembly. Possible causes: MDS link or faulty amplifier bus ribbon cable.

2240255 Y Axis Gradient Amplifier Status. The auxiliary voltage is present but the high voltage is off. Verify the REM/LOC switch on front panel of the System Control Assembly (SCA) in the Gradient Subsystem is set to REM.

2240263 Y-axis gradient amplifier fault 13. AMP No. 1 Fri May 18 11:37:50 2001 Error: 2240263 Host: oc001 Process: NSP File: HPC_PORT_ID:spi_scif Line: 150 Y-axis gradient amplifier fault 13. AMP No. 1 SPI received unused error from GPS.

2240271 Z-axis gradient amplifier state 5. AMP No. 1 Fri May 18 11:38:20 2001 Error: 2240271 Host: oc001 Process: NSP File: HPC_PORT_ID:spi_scif Line: 160 Z-axis gradient amplifier state 5. AMP No. 1 Auxiliary voltage present. This is just for message...

2240239 X-axis gradient amplifier state 5. AMP No. 1 Fri May 18 11:37:50 2001 Error: 2240239 Host: oc001 Process: NSP File: HPC_PORT_ID:spi_scif Line: 140 X-axis gradient amplifier state 5. AMP No. 1 Auxiliary voltage present. This is 

Applies to

Product
Profile 3 Ultrasparc CPU with AN8290Y gradients
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare MR issues