GE Healthcare · MR2240271224023922402632239849
No ID Flt Displayed on the 8290 gradient SCA Board. profile
Symptom
No ID Flt Displayed on the 8290 Gradient SCA Board. The SCA display showed a No ID flt message. The gradients were not ready. X and Z were in stand-by (state 5), Y amp had error 13. Several different errors were reported in the system error log. The problem would occur when powering up.
Error message
2240271 Z-axis gradient amplifier state 5. AMP No. 1 Fri May 18 11:38:20 2001 Error: 2240271 Host: oc001 Process: NSP File: HPC_PORT_ID:spi_scif Line: 160 Z-axis gradient amplifier state 5. AMP No. 1 Auxiliary voltage present. This is just for message... 2240239 X-axis gradient amplifier state 5. AMP No. 1 Fri May 18 11:37:50 2001 Error: 2240239 Host: oc001 Process: NSP File: HPC_PORT_ID:spi_scif Line: 140 X-axis gradient amplifier state 5. AMP No. 1 Auxiliary voltage present. This is just for message... 2240263 Y-axis gradient amplifier fault 13. AMP No. 1 Fri May 18 11:37:50 2001 Error: 2240263 Host: oc001 Process: NSP File: HPC_PORT_ID:spi_scif Line: 150 Y-axis gradient amplifier fault 13. AMP No. 1 SPI received unused error from GPS. 2239849 Gradient Subsystem Clock Loss. Fri May 18 11:41:58 2001 Error: 2239849 Host: oc001 Process: NSP File: HPC_PORT_ID:spi_scif Line: 510 Gradient Subsystem Clock Loss. Verify the clock cable between the TYME II card and the gradient subsystem.
Applies to
- Product
- Profile 3 Ultrasparc CPU with AN8290Y gradients
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Digital Artifact on Axial Lumbar with the Body Flex2 coil.
- MDS Link faults and Loss of AC Mains.
- Gradient Amplifier Fault.
- Analogic 8280 Gradients Fail to Come Up with HV PS Fault Due to Gradient Amplifier
- No ID Flt Displayed on the 8290 Gradient SCA Board.
- No ID Flt Displayed on the 8290 Gradient SCA Board.