No ID Flt Displayed on the 8290 gradient SCA Board. profile

Symptom

No ID Flt Displayed on the 8290 Gradient SCA Board. The SCA display showed a No ID flt message. The gradients were not ready. X and Z were in stand-by (state 5), Y amp had error 13. Several different errors were reported in the system error log. The problem would occur when powering up.

Error message

2240271 Z-axis gradient amplifier state 5. AMP No. 1 Fri May 18 11:38:20 2001 Error: 2240271 Host: oc001 Process: NSP File: HPC_PORT_ID:spi_scif Line: 160 Z-axis gradient amplifier state 5. AMP No. 1 Auxiliary voltage present. This is just for message...

2240239 X-axis gradient amplifier state 5. AMP No. 1 Fri May 18 11:37:50 2001 Error: 2240239 Host: oc001 Process: NSP File: HPC_PORT_ID:spi_scif Line: 140 X-axis gradient amplifier state 5. AMP No. 1 Auxiliary voltage present. This is just for message...

2240263 Y-axis gradient amplifier fault 13. AMP No. 1 Fri May 18 11:37:50 2001 Error: 2240263 Host: oc001 Process: NSP File: HPC_PORT_ID:spi_scif Line: 150 Y-axis gradient amplifier fault 13. AMP No. 1 SPI received unused error from GPS.

2239849 Gradient Subsystem Clock Loss. Fri May 18 11:41:58 2001 Error: 2239849 Host: oc001 Process: NSP File: HPC_PORT_ID:spi_scif Line: 510 Gradient Subsystem Clock Loss. Verify the clock cable between the TYME II card and the gradient subsystem.

Applies to

Product
Profile 3 Ultrasparc CPU with AN8290Y gradients
Area
MR

Full cause & step-by-step resolution