GE Healthcare · MRSigna2240239224025922402432240271

Gradient Amplifier Fault.

Symptom

Will Not Scan with Gradient Errors.

Error message

2240239 X Axis Gradient Amplifier Status. The auxiliary voltage is present but the high voltage is off. Tue Aug 3 12:11:39 2004 Error: 2240239 Host: lx-mr Process: NSP File: HPC_PORT_ID:spi_gca_error Line: 5 X Axis Gradient Amplifier Status. The auxiliary voltage is present but the high voltage is off. Verify the "REM/LOC" switch on front panel of the System Control.

2240259 Gradient Amplifier Fault. Gradient amplifier fault, over current or clock loss internal to Gradient Subsystem Aug 3 Mon 08:20:27 2240259 Y Gradient Amplifier Fault.Gradient amplifier fault, over current or clock loss internal to Gradient Subsystem (GS). Disconnect Y amp bus cable and power on GS in local mode. If GS turns on, replace Y Gradient Amp; else, check amp bus cable or SCA.

2240243 X Gradient Amplifier Fault. Gradient amplifier fault, over current or clock loss internal to Gradient Subsystem See Long Description Fri Jan 28 07:30:43 2005 Error: 2240243 Host: lsdi Process: NSPFile: HPC_PORT_ID:spi_gca_error Line: 9 X Gradient Amplifier Fault. Gradient amplifier fault, over current or clock loss internal to Gradient Subsystem (GS). Disconnect X amp bus cable and power on GS in local mode. If GS turns on

Applies to

Models
Signa
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Software
9.1 For Signa LX product only
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare MR issues