GE Healthcare · MR2224846

The BegTravel and EndTravel parameters in MRconfig.cfg are set incorrectly.

Symptom

ACR Slice Position Accuracy Test Fails.

Error message

2224846 The encoder calibration value was invalid. Please run the cradle calibration routine. Specified Calibration Constant : 6196 Minimum Calibration Constant : 2250 Maximum Calibration Constant : 2750 2224846 pcmr NSP SCP:SRI 100 The encoder calibration value was invalid. Please run the cradle calibration routine. Specified Calibration Constant : 6196 Minimum Calibration Constant : 2250 Maximum Calibration Constant : 2750

Applies to

Product
Open Speed with EXCITE For HFO4 and HFO5 only
Software
OPENMR4-HFO4 Software version for HFO4 system, which is OpenSpeed with Excite Technology
Area
MR

Full cause & step-by-step resolution