Unable to Boot Applications Software due to Software Corruption Caused by Repeated Power Brownouts.

Symptom

Cannot Boot Applications with Host Monitor Timeout Error

Error message

Hostmonitor Timeout after 100 seconds

2248755 RFampType in MRconfig.cfg file does not match RF Amplifier Type determined by SCP. Please update MRconfig.cfg file with correct RFampType and perform a TPS Reset. Actual RF Amplifier Type: 1.5T Solid State SRFD2 2248755 pcmr NSP SCP:RfNbMdsHWO 2325 RFampType in MRconfig.cfg file does not match RF Amplifier Type determined by SCP. Please update MRconfig.cfg file with correct RFampType and perform a TPS Reset. Actual RF Amplifier Type: 1.5T Solid State SRFD2

2224846 The encoder calibration value was invalid. Please run the cradle calibration routine. Specified Calibration Constant : 6196 Minimum Calibration Constant : 2250 Maximum Calibration Constant : 2750 2224846 pcmr NSP SCP:SRI 100 The encoder calibration value was invalid. Please run the cradle calibration routine. Specified Calibration Constant : 6196 Minimum Calibration Constant : 2250 Maximum Calibration Constant : 2750

200002352 A HOST_SHUTDOWN was initiated by the death of sdm. 200002352 pcmr hostMonitor ssProcStop.c 812 A HOST_SHUTDOWN was initiated by the death of sdm.

2248664 - MDS link failure. An MDS device failed to respond to SCP Mon Feb 2 13:27:26 2004 2248664 1 MR-

Applies to

Models
Excite
Product
Excite HD For 12.0 1.5T Systems only.
Software
12.0 Excite HD and 3T Excite HD products only
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.