GE Healthcare · MRExcite22487552224846200002352224866422483492223817
Unable to Boot Applications Software due to Software Corruption Caused by Repeated Power Brownouts.
Symptom
Cannot Boot Applications with Host Monitor Timeout Error
Error message
Hostmonitor Timeout after 100 seconds 2248755 RFampType in MRconfig.cfg file does not match RF Amplifier Type determined by SCP. Please update MRconfig.cfg file with correct RFampType and perform a TPS Reset. Actual RF Amplifier Type: 1.5T Solid State SRFD2 2248755 pcmr NSP SCP:RfNbMdsHWO 2325 RFampType in MRconfig.cfg file does not match RF Amplifier Type determined by SCP. Please update MRconfig.cfg file with correct RFampType and perform a TPS Reset. Actual RF Amplifier Type: 1.5T Solid State SRFD2 2224846 The encoder calibration value was invalid. Please run the cradle calibration routine. Specified Calibration Constant : 6196 Minimum Calibration Constant : 2250 Maximum Calibration Constant : 2750 2224846 pcmr NSP SCP:SRI 100 The encoder calibration value was invalid. Please run the cradle calibration routine. Specified Calibration Constant : 6196 Minimum Calibration Constant : 2250 Maximum Calibration Constant : 2750 200002352 A HOST_SHUTDOWN was initiated by the death of sdm. 200002352 pcmr hostMonitor ssProcStop.c 812 A HOST_SHUTDOWN was initiated by the death of sdm. 2248664 - MDS link failure. An MDS device failed to respond to SCP Mon Feb 2 13:27:26 2004 2248664 1 MR-
Applies to
- Models
- Excite
- Product
- Excite HD For 12.0 1.5T Systems only.
- Software
- 12.0 Excite HD and 3T Excite HD products only
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.