The BegTravel and EndTravel parameters in MRconfig.cfg are set incorrectly.

Symptom

ACR Slice Position Accuracy Test Fails.

Error message

2224846 The encoder calibration value was invalid. Please run the cradle calibration routine. Specified Calibration Constant : 6196 Minimum Calibration Constant : 2250 Maximum Calibration Constant : 2750 2224846 pcmr NSP SCP:SRI 100 The encoder calibration value was invalid. Please run the cradle calibration routine. Specified Calibration Constant : 6196 Minimum Calibration Constant : 2250 Maximum Calibration Constant : 2750

Applies to

Models
Ovation
Product
Ovation with EXCITE
Software
OPENMR4-MFO4 Applies to Ovation4 (Excite) only. Ovation4 (MFO4) is 11.x based software.
Area
MR