Cradle Moves Very Slow in the Longitudinal Direction.

Symptom

Cradle Moves Very Slowly in the Longitudinal Direction See Long Description During longitudinal movement ( in or out), while press and hold "fast in" or "fast out" button, the cradle will start to move very slowly and keep this speed for 50cm. After that the cradle will stop moving. Need to release the button and press and hold it again for the next 50cm of travel. Same thing happens when you press the HOME button. After 50cm the cradle will stop so you need press and hold the Home button again..

Error message

2224846 The encoder calibration value was invalid. Please run the cradle calibration routine. Specified Calibration Constant : 6196 Minimum Calibration Constant : 2250 Maximum Calibration Constant : 2750 2224846 pcmr NSP SCP:SRI 100 The encoder calibration value was invalid. Please run the cradle calibration routine. Specified Calibration Constant : 6196 Minimum Calibration Constant : 2250 Maximum Calibration Constant : 2750

Applies to

Models
Ovation
Product
Ovation with EXCITE
Software
OPENMR4-MFO4 Applies to Ovation4 (Excite) only. Ovation4 (MFO4) is 11.x based software.
Component
2251519 Motor/Shaft Assy Motor and gearbox
Area
MR

Full cause & step-by-step resolution