Cradle Moves Very Slow in the Longitudinal Direction.
Symptom
Cradle Moves Very Slowly in the Longitudinal Direction See Long Description During longitudinal movement ( in or out), while press and hold "fast in" or "fast out" button, the cradle will start to move very slowly and keep this speed for 50cm. After that the cradle will stop moving. Need to release the button and press and hold it again for the next 50cm of travel. Same thing happens when you press the HOME button. After 50cm the cradle will stop so you need press and hold the Home button again..
Error message
2224846 The encoder calibration value was invalid. Please run the cradle calibration routine. Specified Calibration Constant : 6196 Minimum Calibration Constant : 2250 Maximum Calibration Constant : 2750 2224846 pcmr NSP SCP:SRI 100 The encoder calibration value was invalid. Please run the cradle calibration routine. Specified Calibration Constant : 6196 Minimum Calibration Constant : 2250 Maximum Calibration Constant : 2750
Applies to
- Models
- Ovation
- Product
- Ovation with EXCITE
- Software
- OPENMR4-MFO4 Applies to Ovation4 (Excite) only. Ovation4 (MFO4) is 11.x based software.
- Component
- 2251519 Motor/Shaft Assy Motor and gearbox
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA ID100060 Unable To Film.
- CWA ID100315. Auto and Manual Prescan Fails.
- What is the Acoustic Noise Level in the Ovation Magnet.
- Having Difficulties Setting Up Gradcal Scan From Dir 5112585 - Signa Ovation With Excite Technology
- CWA ID101091. Some or All Images in the Last Scan Did Not Reconstruct.
- HFO Cradle Will Not Move Due To Loss Of Home Flag Position.