DQA Scan Completes but will not Calculate.

Symptom

DQA Scan Completes but will not Calculate. Message appears in a pop-up window stating axial slice location is incorrect. Only error associated with cradle movement is 2224846 in the system log. Simply moving the phantom inward 1-2mm, then scanning again, gives a passing result.

Error message

2224846 The encoder calibration value was invalid. Please run the cradle calibration routine. Specified Calibration Constant : 6196 Minimum Calibration Constant : 2250 Maximum Calibration Constant : 2750 2224846 pcmr NSP SCP:SRI 100 The encoder calibration value was invalid. Please run the cradle calibration routine. Specified Calibration Constant : 6196 Minimum Calibration Constant : 2250 Maximum Calibration Constant : 2750

Applies to

Models
Ovation
Product
Ovation with EXCITE
Software
OPENMR4-MFO4 Applies to Ovation4 (Excite) only. Ovation4 (MFO4) is 11.x based software.
Area
MR

Full cause & step-by-step resolution