Troubleshoot starter abnormalities (ST-7011,B / ST-7013) — E1-E8, inverter display codes, shaft lock release
Use this when the starter reports an error after gantry power-up, during anode rotation, or via its microcomputer-circuit / inverter LED displays. This page covers three things the source groups under "Abnormality in the starter": the startup-timing trap that produces a false "Starter setup" error, the full E1-E8 microcomputer error-message table (with its tube-type-mismatch condition), the invert
Tools & parts
| Item | Part number | Qty |
|---|---|---|
| Starter (chassis assembly) — ST-7011,B or ST-7013 | n/a in source | |
| X-ray tube — CXB-400B (as printed in this section; see naming note) | n/a in source | |
| H.E. filter (heat exchanger filter) | n/a in source | |
| 24-V cable, CNS03 | n/a in source |
Full step-by-step for Troubleshoot starter abnormalities (ST-7011,B / ST-7013) — E1-E8, inverter display codes, shaft lock release
All 3 steps are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with the real figures from the source.
Open the full procedureRelated X-ray Generator/HV guides
Source: 2D201-113EN*J §1.11.7 \"Abnormality in the starter\" (pp.123-126), including §1.11.7.1 \"Shaft lock release processing\. Preview re-authored by FSELIB — the full procedure with figures and steps is in the paid service manual.