Check the X-ray tube current with the tube ammeter (MA-1201D)

Measure the X-ray tube current with an external tube ammeter wired into the anode side of the high-voltage circuit, then convert that anode-side reading to the actual cathode-side tube current using the CXB-400C's 0.9 conversion factor.

Applies to
Activion16 (TSX-031A) X-ray generator/HV section, X-ray tube CXB-400C. The 0.9 cathode-side conversion factor given here is specific to the CXB-400C; source states the calculation \"differs depending on the tube used\" without giving another tube's factor in this section.
Steps
4 steps
Source
2D201-113EN*J §1.9.7 'Tube ammeter' (p.64)

Tools & parts

ItemPart numberQty
Tube ammeterMA-1201D (ARCO, or equivalent)1
Tube ammeter detector tankn/a in source1
Torek AB-201515DAB-201515D (TOREK)1

Full step-by-step for Check the X-ray tube current with the tube ammeter (MA-1201D)

All 4 steps are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with the real figures from the source.

Open the full procedure

Related X-ray Generator/HV guides

View all Activion16 (TSX-031A) guides

Source: 2D201-113EN*J §1.9.7 'Tube ammeter' (p.64). Preview re-authored by FSELIB — the full procedure with figures and steps is in the paid service manual.