ProcedureX-ray Generator/HVSetup & CalibrationAdjust
Adjust the If (filament) value remotely from the console
Smooth out the leading edge of the X-ray tube's current waveform by remotely adjusting the If (filament) value, using the <<System maintenance>> console program's "If" entry. The program walks through retrieving the current If value data, selecting which kV/mA/focus conditions to adjust, running automatic warm-up exposures that compare the measured tube current against the stored value, and finall
Applies to
Activion16 (TSX-031A). No model/revision restriction is stated in source for this section. Typically performed after the tube connection/operation check (§1.10.1) and X-ray tube aging (§1.10.2) — following X-ray tube replacement — or whenever the tube-current (mA) waveform observed on the oscilloscope does not match the stored If value.
Steps
31 steps
Source
2D201-113EN*J §1.10.3 'If adjustment' (pp.69-87)
Tools & parts
| Item | Part number | Qty |
|---|---|---|
| Oscilloscope | n/a in source | 1 |
| 3.5-inch floppy disk (FD), MS-DOS format | n/a in source | 1 |
| Navi-port console | n/a in source | 1 |
Full step-by-step for Adjust the If (filament) value remotely from the console
All 31 steps are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with the real figures from the source.
Open the full procedureRelated X-ray Generator/HV guides
Check the X-ray generator power-on and power-off sequence2 stepsCheck the X-ray tube current with the tube ammeter (MA-1201D)4 stepsCheck the X-ray tube voltage with the tube voltage meter (kV-201 bleeder)5 stepsHandle the high-voltage cable — removal, cleaning, and silicone-oil reassembly (CLAYMOUNT)15 stepsReplace the X-ray generator fuse2 stepsRetrieve and decode the XC PWB error log (direct RS-232C read + system-software XL30 console log)11 steps
View all Activion16 (TSX-031A) guidesSource: 2D201-113EN*J §1.10.3 'If adjustment' (pp.69-87). Preview re-authored by FSELIB — the full procedure with figures and steps is in the paid service manual.