Pular para o conteúdo principal
TroubleshootingDASTroubleshootingTroubleshoot

Data acquisition faults — cause × symptom matrix and 4 isolation flowcharts (DAS)

Use this page when a scan shows a ring image, a streak shower, a shifted CT number, or an abnormally large image SD. Source presents this as one summary table — which components can cause which of the four symptoms — followed by a dedicated fault-isolation flowchart for each symptom. This atom carries all of it: the matrix plus all four flowcharts.

Applies to
Activion16 (TSX-031A), CDAS-035A/1A data acquisition section (Converter-16 / Control-16 PWBs). The CDAS-035A/2A 'AJ' generation carries the same matrix and flowcharts at §2.16.3, with AJ-CONVERTER / AJ-CONTROL PWB names in place of Converter-16 / Control-16 — see Related.
Steps
8 steps
Source
2D201-113EN*J §2.8.3 'Possible problems and troubleshooting' (pp.182-186): cause x symptom matrix (p.182) + 4 troubleshooting flowcharts — Ring image (p.183), Streak shower (p.184), Shifted CT number (p.185), SD trouble (p.186)

Tools & parts

ItemPart numberQty
Main detector (body)n/a in source
REF detectorn/a in source
Converter-16 PWBn/a in source
Control-16 PWBn/a in source
DAS power supply / power cablen/a in source
Backplane PWB (main-detector-to-backplane connection)n/a in source
Temperature controllern/a in source
Heatern/a in source
Thermal sensorn/a in source
OPCONTAn/a in source
GCIFAn/a in source
KGTSMn/a in source

Safety

WARNING

Hazard: Working these checks means opening the DAS chassis and, in several branches, powering the gantry/X-ray system to acquire data (cross-referenced from DAS — before repair / general precautions; not restated as new source text in §2.8.3)

Consequence: electric shock or injury from unexpected system operation

Avoidance: turn OFF the breaker and all system power switches at the distribution board before opening the DAS chassis; supply power only to the parts under test while running the DCA acquisitions this page calls for

WARNING

Hazard: Several checks below actually fire the X-ray tube (Check the tube voltage scanning; Check the X-ray output by DCA; Acquire data in DCA/DAS TEST pattern; Is the DAS power supply proper; X-ray tube current OK) — this is not stated as a boxed callout in §2.8.3 itself, but is cross-referenced from the general X-ray-exposure precaution

Consequence: unintended radiation exposure or X-ray leakage from the room

Avoidance: before starting any acquisition in these flowcharts, confirm the door is closed, no persons or X-ray-sensitive materials are in the scanner room, and no X-rays leak from the room

NOTICE

Hazard: Touching a Converter-16 or Control-16 PWB with bare hands while carrying out the repeated 'exchange the PWB with the adjacent/another-slot PWB' steps in these flowcharts (§2.8.1)

Consequence: invisible ESD damage to the board

Avoidance: wear clean cotton gloves whenever a PWB is handled; if a PWB is handled bare-handed, clean it with ethanol

NOTICE

Hazard: Touching the DAS PWB parts or pattern surface with hands or objects while running these checks

Consequence: oil/water contamination degrades signal quality ("SD", as termed in source) — directly relevant to the Large image SD / SD trouble symptom this page diagnoses

Avoidance: avoid touching the DAS PWB parts or pattern surface

NOTICE

Hazard: Selecting the wrong DAS gain when running a DCA acquisition in any of these flowcharts — ordinary image acquisition auto-selects gain, but DCA acquisition requires manual gain selection

Consequence: the DCA output count can read extremely small or overflow, producing a false result on any 'check the X-ray output by DCA' / offset-acquisition step below

Avoidance: for DCA acquisition, manually select the DAS gain appropriate to slice thickness, FOV, and kV before reading the result

NOTICE

Hazard: Main detector temperature outside its stable band while running the Ring-image flowchart's first check

Consequence: image artifacts (ring image)

Avoidance: keep the main detector at the required temperature — this is exactly what the temperature-controller display (36°C to 38°C normal) checked in step 1 of the Ring image flowchart verifies

Full step-by-step for Data acquisition faults — cause × symptom matrix and 4 isolation flowcharts (DAS)

All 8 steps are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with the real figures from the source.

Open the full procedure

Related DAS guides

View all Activion16 (TSX-031A) guides

Source: 2D201-113EN*J §2.8.3 'Possible problems and troubleshooting' (pp.182-186): cause x symptom matrix (p.182) + 4 troubleshooting flowcharts — Ring image (p.183), Streak shower (p.184), Shifted CT number (p.185), SD trouble (p.186). Preview re-authored by FSELIB — the full procedure with figures and steps is in the paid service manual.