Data acquisition faults — cause × symptom matrix and 4 isolation flowcharts (DAS)
Use this page when a scan shows a ring image, a streak shower, a shifted CT number, or an abnormally large image SD. Source presents this as one summary table — which components can cause which of the four symptoms — followed by a dedicated fault-isolation flowchart for each symptom. This atom carries all of it: the matrix plus all four flowcharts.
Tools & parts
| Item | Part number | Qty |
|---|---|---|
| Main detector (body) | n/a in source | |
| REF detector | n/a in source | |
| Converter-16 PWB | n/a in source | |
| Control-16 PWB | n/a in source | |
| DAS power supply / power cable | n/a in source | |
| Backplane PWB (main-detector-to-backplane connection) | n/a in source | |
| Temperature controller | n/a in source | |
| Heater | n/a in source | |
| Thermal sensor | n/a in source | |
| OPCONTA | n/a in source | |
| GCIFA | n/a in source | |
| KGTSM | n/a in source |
Safety
Hazard: Working these checks means opening the DAS chassis and, in several branches, powering the gantry/X-ray system to acquire data (cross-referenced from DAS — before repair / general precautions; not restated as new source text in §2.8.3)
Consequence: electric shock or injury from unexpected system operation
Avoidance: turn OFF the breaker and all system power switches at the distribution board before opening the DAS chassis; supply power only to the parts under test while running the DCA acquisitions this page calls for
Hazard: Several checks below actually fire the X-ray tube (Check the tube voltage scanning; Check the X-ray output by DCA; Acquire data in DCA/DAS TEST pattern; Is the DAS power supply proper; X-ray tube current OK) — this is not stated as a boxed callout in §2.8.3 itself, but is cross-referenced from the general X-ray-exposure precaution
Consequence: unintended radiation exposure or X-ray leakage from the room
Avoidance: before starting any acquisition in these flowcharts, confirm the door is closed, no persons or X-ray-sensitive materials are in the scanner room, and no X-rays leak from the room
Hazard: Touching a Converter-16 or Control-16 PWB with bare hands while carrying out the repeated 'exchange the PWB with the adjacent/another-slot PWB' steps in these flowcharts (§2.8.1)
Consequence: invisible ESD damage to the board
Avoidance: wear clean cotton gloves whenever a PWB is handled; if a PWB is handled bare-handed, clean it with ethanol
Hazard: Touching the DAS PWB parts or pattern surface with hands or objects while running these checks
Consequence: oil/water contamination degrades signal quality ("SD", as termed in source) — directly relevant to the Large image SD / SD trouble symptom this page diagnoses
Avoidance: avoid touching the DAS PWB parts or pattern surface
Hazard: Selecting the wrong DAS gain when running a DCA acquisition in any of these flowcharts — ordinary image acquisition auto-selects gain, but DCA acquisition requires manual gain selection
Consequence: the DCA output count can read extremely small or overflow, producing a false result on any 'check the X-ray output by DCA' / offset-acquisition step below
Avoidance: for DCA acquisition, manually select the DAS gain appropriate to slice thickness, FOV, and kV before reading the result
Hazard: Main detector temperature outside its stable band while running the Ring-image flowchart's first check
Consequence: image artifacts (ring image)
Avoidance: keep the main detector at the required temperature — this is exactly what the temperature-controller display (36°C to 38°C normal) checked in step 1 of the Ring image flowchart verifies
Full step-by-step for Data acquisition faults — cause × symptom matrix and 4 isolation flowcharts (DAS)
All 8 steps are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with the real figures from the source.
Open the full procedureRelated DAS guides
Source: 2D201-113EN*J §2.8.3 'Possible problems and troubleshooting' (pp.182-186): cause x symptom matrix (p.182) + 4 troubleshooting flowcharts — Ring image (p.183), Streak shower (p.184), Shifted CT number (p.185), SD trouble (p.186). Preview re-authored by FSELIB — the full procedure with figures and steps is in the paid service manual.