Check the DAS supply voltage (Control-16 / Converter-16 PWBs)

Verify the DC supply voltages the DAS power supply unit delivers to the Converter-16 and Control-16 PWBs through connectors PCN2 and PCN3, and check the ripple on those same voltages with an oscilloscope. This is a bench check with the power cable and I/O cable already connected — you are reading supply rails at the board connectors, not commanding a scan.

Applies to
Activion16 (TSX-031A), CDAS-035A/1A data acquisition section (Control-16 / Converter-16 PWBs). No model- or revision-specific restriction beyond that generation is stated in source for this task; the CDAS-035A/2A 'AJ' generation has its own byte-identical-values procedure at §2.16.7 (Related), with AJ-CONTROL/AJ-CONVERTER board names in place of Control-16/Converter-16.
Steps
3 steps
Source
2D201-113EN*J §2.8.7 'Check of the DAS supply voltage' (p.189)

Tools & parts

ItemPart numberQty
Digital voltmetern/a in source1
Oscilloscopen/a in source1
Power cablen/a in source1
I/O cablen/a in source1
Converter-16 PWBn/a in source1
Control-16 PWBn/a in source1
V.ADJ controlsn/a in sourcen/a

Full step-by-step for Check the DAS supply voltage (Control-16 / Converter-16 PWBs)

All 3 steps are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with the real figures from the source.

Open the full procedure

Related DAS guides

View all Activion16 (TSX-031A) guides

Source: 2D201-113EN*J §2.8.7 'Check of the DAS supply voltage' (p.189). Preview re-authored by FSELIB — the full procedure with figures and steps is in the paid service manual.