Skip to main content
ProcedureDASFunctional ChecksCheck

Check the DAS supply voltage (Control-16 / Converter-16 PWBs)

Verify the DC supply voltages the DAS power supply unit delivers to the Converter-16 and Control-16 PWBs through connectors PCN2 and PCN3, and check the ripple on those same voltages with an oscilloscope. This is a bench check with the power cable and I/O cable already connected — you are reading supply rails at the board connectors, not commanding a scan.

Applies to
Activion16 (TSX-031A), CDAS-035A/1A data acquisition section (Control-16 / Converter-16 PWBs). No model- or revision-specific restriction beyond that generation is stated in source for this task; the CDAS-035A/2A 'AJ' generation has its own byte-identical-values procedure at §2.16.7 (Related), with AJ-CONTROL/AJ-CONVERTER board names in place of Control-16/Converter-16.
Steps
3 steps
Source
2D201-113EN*J §2.8.7 'Check of the DAS supply voltage' (p.189)

Tools & parts

ItemPart numberQty
Digital voltmetern/a in source1
Oscilloscopen/a in source1
Power cablen/a in source1
I/O cablen/a in source1
Converter-16 PWBn/a in source1
Control-16 PWBn/a in source1
V.ADJ controlsn/a in sourcen/a

Full step-by-step for Check the DAS supply voltage (Control-16 / Converter-16 PWBs)

All 3 steps are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with the real figures from the source.

7-day free trial · no credit card required

Related DAS guides

View all Activion16 (TSX-031A) guides

Related error codes

Source: 2D201-113EN*J §2.8.7 'Check of the DAS supply voltage' (p.189). Preview re-authored by FSELIB — the full procedure with figures and steps is in the paid service manual.

“That they may see, and know, and consider, and understand together, that the hand of the LORD hath done this.”Isaiah 41:20