Jack up the patient couch when it cannot move up from the DOWN limit (emergency)
Use this when the patient couch is disabled and cannot move up from the DOWN limit — you must raise it mechanically with an automotive jack so you can check the parts inside the couch base.
Tools & parts
| Item | Part number | Qty |
|---|---|---|
| Automotive/car jack | n/a in source | 1 |
| Load-spreading platform, 90 cm x 90 cm | n/a in source | 1 |
| Square piece of wood | n/a in source | 1 |
Safety
Hazard: Inspecting the patient couch while it is raised on an automotive jack
Consequence: The upper frame may fall if the jack slips, causing injury
Avoidance: Place a square piece of wood between the frame and the floor before working under/around the raised couch, so the frame is held up even if the jack slips
Full step-by-step for Jack up the patient couch when it cannot move up from the DOWN limit (emergency)
All 3 steps are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with the real figures from the source.
Open the full procedureRelated Couch/Table guides
Source: 2D201-112EN*D §2.4.1 (p.132-133) — Jacking up the patient couch. Preview re-authored by FSELIB — the full procedure with figures and steps is in the paid service manual.