Pular para o conteúdo principal
ProcedureCouch/TableTroubleshootingEmergency

Jack up the patient couch when it cannot move up from the DOWN limit (emergency)

Use this when the patient couch is disabled and cannot move up from the DOWN limit — you must raise it mechanically with an automotive jack so you can check the parts inside the couch base.

Applies to
Activion16 (TSX-031A). Emergency-only workaround for when the patient couch cannot move up from the DOWN limit; no couch-variant restriction is noted in the source for this task.
Steps
3 steps
Source
2D201-112EN*D §2.4.1 (p.132-133) — Jacking up the patient couch

Tools & parts

ItemPart numberQty
Automotive/car jackn/a in source1
Load-spreading platform, 90 cm x 90 cmn/a in source1
Square piece of woodn/a in source1

Safety

CAUTION

Hazard: Inspecting the patient couch while it is raised on an automotive jack

Consequence: The upper frame may fall if the jack slips, causing injury

Avoidance: Place a square piece of wood between the frame and the floor before working under/around the raised couch, so the frame is held up even if the jack slips

Full step-by-step for Jack up the patient couch when it cannot move up from the DOWN limit (emergency)

All 3 steps are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with the real figures from the source.

Open the full procedure

Related Couch/Table guides

View all Activion16 (TSX-031A) guides

Source: 2D201-112EN*D §2.4.1 (p.132-133) — Jacking up the patient couch. Preview re-authored by FSELIB — the full procedure with figures and steps is in the paid service manual.