ProcedureCouch/TableSetup & CalibrationAdjust
Adjust the couch height display
Calibrate the couch-height data shown on the operator panel so it reads 0 at the HEIGHT-1 couch position and +175 at the UP LIMIT position, by trimming two potentiometers on the KGTSM PWB.
Applies to
Activion16 (TSX-031A). No model/revision restriction is stated for this section.
Steps
6 steps
Source
2D201-112EN*D §3.8.7 (p.294) — Adjustment of couch height displays
Tools & parts
| Item | Part number | Qty |
|---|---|---|
| Screwdriver for adjustment | n/a in source | |
| KGTSM PWB — DIP switch DSW4, bit 4 | n/a in source | |
| KGTSM PWB — VR6 (HEIG/OFFSET) | n/a in source | |
| KGTSM PWB — VR7 (HEIG/GAIN) | n/a in source |
Full step-by-step for Adjust the couch height display
All 6 steps are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with the real figures from the source.
Open the full procedureRelated Couch/Table guides
Adjust the couch descent (lowering) speed — solenoid throttle valves L21/L245 stepsAdjust the couch-top longitudinal belt tension6 stepsAdjust the couch-top supporting rollers (height, level, and sway)8 stepsChange the patient couch vertical stop positions2 stepsCheck the couch-top (horizontal) travel speed4 stepsCouch — routing of the internal cables at the linkage unit (work procedure)8 steps
View all Activion16 (TSX-031A) guidesSource: 2D201-112EN*D §3.8.7 (p.294) — Adjustment of couch height displays. Preview re-authored by FSELIB — the full procedure with figures and steps is in the paid service manual.