Pular para o conteúdo principal
ProcedureCouch/TableFunctional ChecksMeasure

Check the couch-top (horizontal) travel speed

Verify the patient couch top's horizontal (IN/OUT) travel speed at its two operator-panel-selectable speeds — 10 mm/s and 130 mm/s — by timing a measured length of travel with a stopwatch and comparing the elapsed time against the source's pass/fail limits.

Applies to
Activion16 (TSX-031A). Couch-top horizontal (longitudinal, IN/OUT) travel at the two selectable operator-panel speeds, 10 mm/s and 130 mm/s.
Steps
4 steps
Source
2D201-112EN*D §3.8.8 (p.295) — Checking the couch top speed

Safety

CAUTION

Hazard: Attaching the cable-tie marker and the measuring tape while the couch top sits farther inward than the OUT2 position

Consequence: Not stated beyond the caution itself — source gives no explicit injury/equipment-damage outcome for this condition

Avoidance: Be careful when attaching the cable tie and the measuring tape whenever the couch top is inward of OUT2

Full step-by-step for Check the couch-top (horizontal) travel speed

All 4 steps are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with the real figures from the source.

Open the full procedure

Related Couch/Table guides

View all Activion16 (TSX-031A) guides

Source: 2D201-112EN*D §3.8.8 (p.295) — Checking the couch top speed. Preview re-authored by FSELIB — the full procedure with figures and steps is in the paid service manual.