Skip to main content
ProcedureCouch/TableSetup & CalibrationAdjust

Change the patient couch vertical stop positions

Fine-tune where the patient couch top's vertical (height) travel is stopped, by repositioning the height-detection photosensor for the stop position you need to change. These stop positions exist for safety — they are what keeps the couch top from interfering with a tilted gantry, from being driven in while already too low, or from over-traveling past its UP/DOWN limits.

Applies to
Activion16 (TSX-031A). No model/revision restriction is called out in this section. Source explicitly scopes this task: perform only per instructions from the factory, and only as a fine adjustment.
Steps
2 steps
Source
2D201-112EN*D §2.6.2 (pp.156–157) — Changing the vertical stop positions

Tools & parts

ItemPart numberQty
PHC21 photosensor (Upper Limit / UP)n/a in source
PHC27 photosensor (HEIGHT-0)n/a in source
PHC26 photosensor (HEIGHT-0.5)n/a in source
PHC23 photosensor (HEIGHT-1)n/a in source
PHC24 photosensor (HEIGHT-2)n/a in source
PHC25 photosensor (Auto-Stop)n/a in source
PHC22 photosensor (Lower Limit / DOWN)n/a in source
Photosensor fixing screwn/a in source

Safety

CAUTION

Hazard: Adjusting the vertical-stop photosensor positions other than as instructed by the factory for the specific unit

Consequence: The stop positions no longer perform their safety function (preventing couch-top/gantry interference and over-travel at the UP/DOWN limits), and the procedure itself calls for only fine adjustment, not a large change

Avoidance: Perform this adjustment only according to instructions from the factory; do not attempt a large repositioning of any photosensor

NOTICE

Hazard: Moving a height-detection photosensor in the direction opposite the one indicated for the stop position being adjusted

Consequence: "Trouble will occur" (source's own wording — a fault in the affected stop position)

Avoidance: Loosen only the fixing screw of the selected sensor and move it in the indicated direction; never move it the opposite way

Full step-by-step for Change the patient couch vertical stop positions

All 2 steps are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with the real figures from the source.

7-day free trial · no credit card required

Related Couch/Table guides

View all Activion16 (TSX-031A) guides

Related error codes

Source: 2D201-112EN*D §2.6.2 (pp.156–157) — Changing the vertical stop positions. Preview re-authored by FSELIB — the full procedure with figures and steps is in the paid service manual.

“That they may see, and know, and consider, and understand together, that the hand of the LORD hath done this.”Isaiah 41:20