N1A — Stator Control Board FRAM verify/error
Philips CT error code, reported by the Stator Control Board (Gantry Control Electronics).
What N1A means
On Access Series CT (6/16/Dual) CT systems, status code N1A indicates: Stator Control Board FRAM verify/error. The code originates in the Gantry Control (Stator Control Board) section of the Stator Control Board (Gantry Control Electronics). Severity level warning (Warning — recoverable condition).
Reported action: This indicates an FRAM (non-volatile memory) verification failure on the Stator Control Board; power-cycle and retry, and replace the board if the warning persists..
Listed in Table 1-3 'Stator Control Board Error Code vs System Error Code' of the Philips Access CT Service Trouble Shooting Guide (459800863401 Rev. C), for the Access Series CT (6/16/Dual) family. No System Error Code mapping is given by the source for this code. Not confirmed for other Philips CT systems.