ER0E — Gantry foot switch error
Philips CT error code, reported by the Stator Control Board (Gantry Control Electronics).
What ER0E means
On Access Series CT (6/16/Dual) CT systems, status code ER0E indicates: Gantry foot switch error. The code originates in the Gantry Control (Stator Control Board) section of the Stator Control Board (Gantry Control Electronics). Severity level alarm (Alarm — stator control board error condition).
Reported action: Cross-reference System Error Code 0x5601 in the system log. Inspect the gantry foot switch, its wiring, and connector at the Stator Control Board before replacing the switch or board..
Listed in Table 1-3 'Stator Control Board Error Code vs System Error Code' of the Philips Access CT Service Trouble Shooting Guide (459800863401 Rev. C), for the Access Series CT (6/16/Dual) family. Not confirmed for other Philips CT systems.
Full troubleshooting for ER0E
The complete diagnosis, cause tree, and step-by-step remedy for the Gantry Control (Stator Control Board) subsystem are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with real figures.
Open the Activion16 service manualRelated Gantry Control (Stator Control Board) codes
Source: Philips Access CT Service Trouble Shooting Guide — 459800863401 Rev. C. Reference information re-authored by FSELIB — not a copy of the OEM manual.