Alarm — stator control board error conditionGantry Control (Stator Control Board)

ER15Stator Control Board A/D converter self-detect failed

Philips CT error code, reported by the Stator Control Board (Gantry Control Electronics).

What ER15 means

On Access Series CT (6/16/Dual) CT systems, status code ER15 indicates: Stator Control Board A/D converter self-detect failed. The code originates in the Gantry Control (Stator Control Board) section of the Stator Control Board (Gantry Control Electronics). Severity level alarm (Alarm — stator control board error condition).

Reported action: Cross-reference System Error Code 0x5608 in the system log. This is an onboard self-test failure of the Stator Control Board's A/D converter; if it repeats after a power cycle, replace the Stator Control Board..

Listed in Table 1-3 'Stator Control Board Error Code vs System Error Code' of the Philips Access CT Service Trouble Shooting Guide (459800863401 Rev. C), for the Access Series CT (6/16/Dual) family. Not confirmed for other Philips CT systems.

Full troubleshooting for ER15

The complete diagnosis, cause tree, and step-by-step remedy for the Gantry Control (Stator Control Board) subsystem are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with real figures.

Open the Activion16 service manual

Related Gantry Control (Stator Control Board) codes

Discuss ER15 with other field engineers in the FSELIB forum

Source: Philips Access CT Service Trouble Shooting Guide — 459800863401 Rev. C. Reference information re-authored by FSELIB — not a copy of the OEM manual.