Alarm — stator control board error conditionGantry Control (Stator Control Board)

ER17CAN inquiry command time-out

Philips CT error code, reported by the Stator Control Board (Gantry Control Electronics).

What ER17 means

On Access Series CT (6/16/Dual) CT systems, status code ER17 indicates: CAN inquiry command time-out. The code originates in the Gantry Control (Stator Control Board) section of the Stator Control Board (Gantry Control Electronics). Severity level alarm (Alarm — stator control board error condition).

Reported action: Cross-reference System Error Code 0x5702 in the system log. Check CAN bus wiring/termination and the addressed CAN node (see Table 1-6 Unit CAN ID) before replacing the Stator Control Board..

Listed in Table 1-3 'Stator Control Board Error Code vs System Error Code' of the Philips Access CT Service Trouble Shooting Guide (459800863401 Rev. C), for the Access Series CT (6/16/Dual) family. Not confirmed for other Philips CT systems.