ER0F — Host-to-Gantry interface board cable abnormal
Philips CT error code, reported by the Stator Control Board (Gantry Control Electronics).
What ER0F means
On Access Series CT (6/16/Dual) CT systems, status code ER0F indicates: Host-to-Gantry interface board cable abnormal. The code originates in the Gantry Control (Stator Control Board) section of the Stator Control Board (Gantry Control Electronics). Severity level alarm (Alarm — stator control board error condition).
Reported action: Cross-reference System Error Code 0x5602 in the system log. Inspect the cable and connectors between the Host computer and the Gantry interface board for damage or a loose connection before replacing the cable or interface board..
Listed in Table 1-3 'Stator Control Board Error Code vs System Error Code' of the Philips Access CT Service Trouble Shooting Guide (459800863401 Rev. C), for the Access Series CT (6/16/Dual) family. Not confirmed for other Philips CT systems. Source text names the far end as 'Ghost' — likely a print/OCR artifact for 'Host'.
Full troubleshooting for ER0F
The complete diagnosis, cause tree, and step-by-step remedy for the Gantry Control (Stator Control Board) subsystem are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with real figures.
Open the Activion16 service manualRelated Gantry Control (Stator Control Board) codes