Pular para o conteúdo principal
Alarm — stator control board error conditionGantry Control (Stator Control Board)

ER0FHost-to-Gantry interface board cable abnormal

Philips CT error code, reported by the Stator Control Board (Gantry Control Electronics).

What ER0F means

On Access Series CT (6/16/Dual) CT systems, status code ER0F indicates: Host-to-Gantry interface board cable abnormal. The code originates in the Gantry Control (Stator Control Board) section of the Stator Control Board (Gantry Control Electronics). Severity level alarm (Alarm — stator control board error condition).

Reported action: Cross-reference System Error Code 0x5602 in the system log. Inspect the cable and connectors between the Host computer and the Gantry interface board for damage or a loose connection before replacing the cable or interface board..

Listed in Table 1-3 'Stator Control Board Error Code vs System Error Code' of the Philips Access CT Service Trouble Shooting Guide (459800863401 Rev. C), for the Access Series CT (6/16/Dual) family. Not confirmed for other Philips CT systems. Source text names the far end as 'Ghost' — likely a print/OCR artifact for 'Host'.

Full troubleshooting for ER0F

The complete diagnosis, cause tree, and step-by-step remedy for the Gantry Control (Stator Control Board) subsystem are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with real figures.

Open the Activion16 service manual

Related Gantry Control (Stator Control Board) codes

Discuss ER0F with other field engineers in the FSELIB forum

Source: Philips Access CT Service Trouble Shooting Guide — 459800863401 Rev. C. Reference information re-authored by FSELIB — not a copy of the OEM manual.