Alarm — stator control board error conditionGantry Control (Stator Control Board)

ER13 — Temperature measurement board communication CRC error

Philips CT error code, reported by the Stator Control Board (Gantry Control Electronics).

What ER13 means

On Access Series CT (6/16/Dual) CT systems, status code ER13 indicates: Temperature measurement board communication CRC error. The code originates in the Gantry Control (Stator Control Board) section of the Stator Control Board (Gantry Control Electronics). Severity level alarm (Alarm — stator control board error condition).

Reported action: Cross-reference System Error Code 0x5606 in the system log. Check for electrical noise or a faulty cable/connector on the temperature measurement board link before replacing either board..

Listed in Table 1-3 'Stator Control Board Error Code vs System Error Code' of the Philips Access CT Service Trouble Shooting Guide (459800863401 Rev. C), for the Access Series CT (6/16/Dual) family. Not confirmed for other Philips CT systems.