Warning — recoverable conditionGantry Control (Stator Control Board)

N18Monitor channel A warning

Philips CT error code, reported by the Stator Control Board (Gantry Control Electronics).

What N18 means

On Access Series CT (6/16/Dual) CT systems, status code N18 indicates: Monitor channel A warning. The code originates in the Gantry Control (Stator Control Board) section of the Stator Control Board (Gantry Control Electronics). Severity level warning (Warning — recoverable condition).

Reported action: Cross-reference System Error Code 0x5728. Monitor for recurrence on monitor channel A; if it repeats together with the channel B warning (n19) or escalates to Er3c, treat as a Stator Control Board issue..

Listed in Table 1-3 'Stator Control Board Error Code vs System Error Code' of the Philips Access CT Service Trouble Shooting Guide (459800863401 Rev. C), for the Access Series CT (6/16/Dual) family. Not confirmed for other Philips CT systems.