N0A — Stator Control Board ComOK signal overtime
Philips CT error code, reported by the Stator Control Board (Gantry Control Electronics).
What N0A means
On Access Series CT (6/16/Dual) CT systems, status code N0A indicates: Stator Control Board ComOK signal overtime. The code originates in the Gantry Control (Stator Control Board) section of the Stator Control Board (Gantry Control Electronics). Severity level warning (Warning — recoverable condition).
Reported action: Cross-reference System Error Code 0x5723. Check the ComOK acknowledgement signal path described in the Data Path Resend Mechanism (Section 2.6.1) between the Rotor Control Board and the DRC board..
Listed in Table 1-3 'Stator Control Board Error Code vs System Error Code' of the Philips Access CT Service Trouble Shooting Guide (459800863401 Rev. C), for the Access Series CT (6/16/Dual) family, in the 'nxx' warning-code range. Not confirmed for other Philips CT systems.
Full troubleshooting for N0A
The complete diagnosis, cause tree, and step-by-step remedy for the Gantry Control (Stator Control Board) subsystem are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with real figures.
Open the Activion16 service manualRelated Gantry Control (Stator Control Board) codes
Source: Philips Access CT Service Trouble Shooting Guide — 459800863401 Rev. C. Reference information re-authored by FSELIB — not a copy of the OEM manual.