Warning — recoverable conditionGantry Control (Stator Control Board)

N07CT scan room door open

Philips CT error code, reported by the Stator Control Board (Gantry Control Electronics).

What N07 means

On Access Series CT (6/16/Dual) CT systems, status code N07 indicates: CT scan room door open. The code originates in the Gantry Control (Stator Control Board) section of the Stator Control Board (Gantry Control Electronics). Severity level warning (Warning — recoverable condition).

Reported action: Close and properly latch the CT room door; check the door interlock switch and wiring if the warning (System Error Code Error0429) persists with the door closed..

Listed in Table 1-3 'Stator Control Board Error Code vs System Error Code' of the Philips Access CT Service Trouble Shooting Guide (459800863401 Rev. C), for the Access Series CT (6/16/Dual) family, in the 'nxx' warning-code range; corresponding System Error Code is given as 'Error0429'. Not confirmed for other Philips CT systems.