Pular para o conteúdo principal
Alarm — stator control board error conditionGantry Control (Stator Control Board)

ER3FStator Control Board write error

Philips CT error code, reported by the Stator Control Board (Gantry Control Electronics).

What ER3F means

On Access Series CT (6/16/Dual) CT systems, status code ER3F indicates: Stator Control Board write error. The code originates in the Gantry Control (Stator Control Board) section of the Stator Control Board (Gantry Control Electronics). Severity level alarm (Alarm — stator control board error condition).

Reported action: This indicates a failed write operation on the Stator Control Board (e.g. EEPROM/FRAM); power-cycle the gantry and retry, and replace the Stator Control Board if the code repeats..

Listed in Table 1-3 'Stator Control Board Error Code vs System Error Code' of the Philips Access CT Service Trouble Shooting Guide (459800863401 Rev. C), for the Access Series CT (6/16/Dual) family. No System Error Code mapping is given by the source for this code. Not confirmed for other Philips CT systems.

Full troubleshooting for ER3F

The complete diagnosis, cause tree, and step-by-step remedy for the Gantry Control (Stator Control Board) subsystem are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with real figures.

Open the Activion16 service manual

Related Gantry Control (Stator Control Board) codes

Discuss ER3F with other field engineers in the FSELIB forum

Source: Philips Access CT Service Trouble Shooting Guide — 459800863401 Rev. C. Reference information re-authored by FSELIB — not a copy of the OEM manual.