Pular para o conteúdo principal
Alarm — stator control board error conditionGantry Control (Stator Control Board)

ER3DStator Control Board CAN cable error

Philips CT error code, reported by the Stator Control Board (Gantry Control Electronics).

What ER3D means

On Access Series CT (6/16/Dual) CT systems, status code ER3D indicates: Stator Control Board CAN cable error. The code originates in the Gantry Control (Stator Control Board) section of the Stator Control Board (Gantry Control Electronics). Severity level alarm (Alarm — stator control board error condition).

Reported action: Inspect the CAN bus cable and connectors at the Stator Control Board for damage or a loose connection; check bus termination before replacing the board..

Listed in Table 1-3 'Stator Control Board Error Code vs System Error Code' of the Philips Access CT Service Trouble Shooting Guide (459800863401 Rev. C), for the Access Series CT (6/16/Dual) family. No System Error Code mapping is given by the source for this code. Not confirmed for other Philips CT systems.

Full troubleshooting for ER3D

The complete diagnosis, cause tree, and step-by-step remedy for the Gantry Control (Stator Control Board) subsystem are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with real figures.

Open the Activion16 service manual

Related Gantry Control (Stator Control Board) codes

Discuss ER3D with other field engineers in the FSELIB forum

Source: Philips Access CT Service Trouble Shooting Guide — 459800863401 Rev. C. Reference information re-authored by FSELIB — not a copy of the OEM manual.