ER36 — Collimator error (grouped code)
Philips CT error code, reported by the Collimator.
What ER36 means
On Access Series CT (6/16/Dual) CT systems, status code ER36 indicates: Collimator error (grouped code). The code originates in the Collimator section of the Collimator. Severity level alarm (Alarm — stator control board error condition).
Reported action: Identify the specific System Error Code (0x6001-0x6026 range) in the log, then follow the Collimator Prepare Error flow chart (Section 2.4.7) and inspect the collimator motor/position feedback..
Listed in Table 1-3 'Stator Control Board Error Code vs System Error Code' of the Philips Access CT Service Trouble Shooting Guide (459800863401 Rev. C), for the Access Series CT (6/16/Dual) family; maps to System Error Code range 0x6001-0x6015 and 0x6019-0x6026. Not confirmed for other Philips CT systems.
Full troubleshooting for ER36
The complete diagnosis, cause tree, and step-by-step remedy for the Collimator subsystem are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with real figures.
Open the Activion16 service manualRelated Collimator codes