200011D6 — Machine powered down abnormally
Philips CT error code, reported by the Power.
What 200011D6 means
On CT systems, status code 200011D6 indicates: Machine powered down abnormally. The code originates in the Power section of the Power. Severity level warning (Abnormal shutdown recorded).
Reported action: Look for the cause of the unexpected power loss: facility power, the mains cord and connections, the system power supply, and the UPS/battery if one is fitted. This code is also useful as a time marker — when file system or database errors appear immediately after it, check whether they are consequences of that power loss before treating them as independent faults..
From the Philips ultrasound service reference "Error Logger Overview" (New Error Logger), archived under Philips ultrasound End-of-Life service documentation. The source lists this Errno as "Machine abnormally powered down" — the system lost power outside a normal shutdown. A full log entry combines a 4-digit Subsystem code with the Errno (format ssss-eeeeeeee).
Full troubleshooting for 200011D6
The complete diagnosis, cause tree, and step-by-step remedy for the Power subsystem are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with real figures.
Open the Activion16 service manualRelated Power codes
Source: Philips Ultrasound (End of Life) — 10_Error Logger Overview, Reference documentation. Reference information re-authored by FSELIB — not a copy of the OEM manual.