Alarm — high severityMUDAT (data transfer system)

B7300000Data transfer between the control PWBs in the gantry is abnormal

Canon Medical (Toshiba) CT error code, reported by the MUDAT (Multi-Data Transmitting System) — optical data-transfer link.

What B7300000 means

On Aquilion RX and Aquilion ONE CT systems, status code B7300000 indicates: Data transfer between the control PWBs in the gantry is abnormal. The code originates in the MUDAT (data transfer system) section of the MUDAT (Multi-Data Transmitting System) — optical data-transfer link. Severity level A1 (Alarm — high severity).

Reported action: Trace the control chain around the KGTSM, which is the hub: confirm it is receiving position and status information from the actuators and from the OPCONTA, and that its control signals are reaching the KGTSP. Check power and connectors on the gantry control PWBs (KGTSM, KGTSP, GCIFA/GCIFA-D, OPCONTA) and inspect the cabling between the rotation and stationary sections, then re-run the gantry operation check before replacing a board..

Toshiba / Canon Medical Aquilion RX and Aquilion ONE CT systems, on the gantry/couch control section. On Aquilion-family systems that section is built from a small set of PWBs — the KGTSM as the central controller, the KGTSP power PWB, the GCIFA and GCIFA-D, the OPCONTA, the SI PWB, the DD motor controller and the operating panels. The KGTSM checks operation requests and gantry/couch status, asks the OPCONTA to operate the rotating unit, and receives rotor status and position information back, so a break in that exchange stops gantry control.

Full troubleshooting for B7300000

The complete diagnosis, cause tree, and step-by-step remedy for the MUDAT (data transfer system) subsystem are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with real figures.