Pular para o conteúdo principal
Alarm — high severityMUDAT (data transfer system)

B7100006Error in MUDAT — generic gantry optical data-transfer fault

Canon Medical (Toshiba) CT error code, reported by the MUDAT (Multi-Data Transmitting System) — optical data-transfer link.

What B7100006 means

On Aquilion RX and Aquilion ONE and Aquilion S16 and Aquilion 16 and Asteion Super4 CT systems, status code B7100006 indicates: Error in MUDAT — generic gantry optical data-transfer fault. The code originates in the MUDAT (data transfer system) section of the MUDAT (Multi-Data Transmitting System) — optical data-transfer link. Severity level A1 (Alarm — high severity).

Reported action: Because the message names no sub-code, start broad and narrow with the six error-identification tools: DAS test and Console test to establish whether the data source or the console side is healthy, then ROTSOT and SOTROT tests for each direction of the link, then OPCONTA test and GCIFA test for the boards. Read the GCIFA and OPCONTA fault LEDs against the manual's data-transfer error table — the LEDs usually name the specific error type the console message omits..

Toshiba / Canon Medical CT systems using the MUDAT optical link (Aquilion RX, Aquilion ONE, Aquilion S16, Aquilion 16 and Asteion Super4). The console message carries no sub-code, so it identifies only the unit: MUDAT, the optical data-transmission unit in the gantry, with the OPCONTA PWB as the rotation-section interface and the GCIFA PWB as the stationary-section interface.

Full troubleshooting for B7100006

The complete diagnosis, cause tree, and step-by-step remedy for the MUDAT (data transfer system) subsystem are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with real figures.

Open the Activion16 service manual

Related MUDAT (data transfer system) codes

Related service procedures

Discuss B7100006 with other field engineers in the FSELIB forum

Source: Toshiba/Canon CT service error message table (errmsg.eng); Aquilion 16 service manual, MUDAT / IF-unit chapter (2D201-059E). Reference information re-authored by FSELIB — not a copy of the OEM manual.