Pular para o conteúdo principal
Alarm — high severityMUDAT (data transfer system)

B710000EError in MUDAT (14) — indexed gantry data-transfer fault

Canon Medical (Toshiba) CT error code, reported by the MUDAT (Multi-Data Transmitting System) — optical data-transfer link.

What B710000E means

On Aquilion RX and Aquilion ONE and Aquilion S16 and Aquilion 16 and Asteion Super4 CT systems, status code B710000E indicates: Error in MUDAT (14) — indexed gantry data-transfer fault. The code originates in the MUDAT (data transfer system) section of the MUDAT (Multi-Data Transmitting System) — optical data-transfer link. Severity level A1 (Alarm — high severity).

Reported action: Identify the fault from the hardware rather than the text: read the GCIFA and OPCONTA fault LEDs and match them against the manual's data-transfer error table, which names the error type and tells you whether the system latches it. Then place the failure with the built-in tools — Console test and DAS test to bound the problem, ROTSOT and SOTROT tests for the transfer direction, OPCONTA and GCIFA tests for the boards..

Toshiba / Canon Medical CT systems using the MUDAT optical link (Aquilion RX, Aquilion ONE, Aquilion S16, Aquilion 16 and Asteion Super4). The message names the MUDAT unit and an index without an error name, so it should be worked like the other MUDAT link faults: from the interface boards outward.

Full troubleshooting for B710000E

The complete diagnosis, cause tree, and step-by-step remedy for the MUDAT (data transfer system) subsystem are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with real figures.

Open the Activion16 service manual

Related MUDAT (data transfer system) codes

Related service procedures

Discuss B710000E with other field engineers in the FSELIB forum

Source: Toshiba/Canon CT service error message table (errmsg.eng); Aquilion 16 service manual, MUDAT / IF-unit chapter (2D201-059E). Reference information re-authored by FSELIB — not a copy of the OEM manual.