Pular para o conteúdo principal
Alarm — high severityMUDAT (data transfer system)

B710000DError in MUDAT (ROTSOT -13) — rotation-to-stationary transfer error

Canon Medical (Toshiba) CT error code, reported by the MUDAT (Multi-Data Transmitting System) — optical data-transfer link.

What B710000D means

On Aquilion RX and Aquilion ONE and Aquilion S16 and Aquilion 16 and Asteion Super4 CT systems, status code B710000D indicates: Error in MUDAT (ROTSOT -13) — rotation-to-stationary transfer error. The code originates in the MUDAT (data transfer system) section of the MUDAT (Multi-Data Transmitting System) — optical data-transfer link. Severity level A1 (Alarm — high severity).

Reported action: Check whether the scan aborted or only logged, because on this direction the more serious error types are latched by the KGTSM while the milder ones are not. Read the GCIFA and OPCONTA fault LEDs, identify the error type in the data-transfer error table together with its documented clearing method, then run the ROTSOT test and, if it passes, the OPCONTA and GCIFA tests to find the board at fault..

Toshiba / Canon Medical CT systems using the MUDAT optical link (Aquilion RX, Aquilion ONE, Aquilion S16, Aquilion 16 and Asteion Super4). ROTSOT covers the rotation-to-stationary direction, which is the direction that carries the acquired raw data; the KGTSM latches the fatal error types on that direction and puts the message on the console screen.

Full troubleshooting for B710000D

The complete diagnosis, cause tree, and step-by-step remedy for the MUDAT (data transfer system) subsystem are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with real figures.

Open the Activion16 service manual

Related MUDAT (data transfer system) codes

Related service procedures

Discuss B710000D with other field engineers in the FSELIB forum

Source: Toshiba/Canon CT service error message table (errmsg.eng); Aquilion 16 service manual, MUDAT / IF-unit chapter (2D201-059E). Reference information re-authored by FSELIB — not a copy of the OEM manual.