B710000C — Error in MUDAT (12) — indexed gantry data-transfer fault
Canon Medical (Toshiba) CT error code, reported by the MUDAT (Multi-Data Transmitting System) — optical data-transfer link.
What B710000C means
On Aquilion RX and Aquilion ONE and Aquilion S16 and Aquilion 16 and Asteion Super4 CT systems, status code B710000C indicates: Error in MUDAT (12) — indexed gantry data-transfer fault. The code originates in the MUDAT (data transfer system) section of the MUDAT (Multi-Data Transmitting System) — optical data-transfer link. Severity level A1 (Alarm — high severity).
Reported action: Capture the gantry-side state before clearing anything: read the fault LEDs on the GCIFA (bottom rear of the gantry) and on the OPCONTA (rotation-section chassis), then map them to the error table to get the real error type, its handling and its clearing method. Follow with the ROTSOT/SOTROT tests for direction and the OPCONTA/GCIFA tests for the boards..
Toshiba / Canon Medical CT systems using the MUDAT optical link (Aquilion RX, Aquilion ONE, Aquilion S16, Aquilion 16 and Asteion Super4). Like the other indexed MUDAT messages, this one names the unit and an index only; the physical detail sits in the GCIFA and OPCONTA LED displays and in the manual's data-transfer error table.
Full troubleshooting for B710000C
The complete diagnosis, cause tree, and step-by-step remedy for the MUDAT (data transfer system) subsystem are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with real figures.
Open the Activion16 service manualRelated MUDAT (data transfer system) codes
Related service procedures
Source: Toshiba/Canon CT service error message table (errmsg.eng); Aquilion 16 service manual, MUDAT / IF-unit chapter (2D201-059E). Reference information re-authored by FSELIB — not a copy of the OEM manual.