B710000B — Error in MUDAT (ROTSOT -11) — rotation-to-stationary transfer error
Canon Medical (Toshiba) CT error code, reported by the MUDAT (Multi-Data Transmitting System) — optical data-transfer link.
What B710000B means
On Aquilion RX and Aquilion ONE and Aquilion S16 and Aquilion 16 and Asteion Super4 CT systems, status code B710000B indicates: Error in MUDAT (ROTSOT -11) — rotation-to-stationary transfer error. The code originates in the MUDAT (data transfer system) section of the MUDAT (Multi-Data Transmitting System) — optical data-transfer link. Severity level A1 (Alarm — high severity).
Reported action: Check the GCIFA first, since it is the board that both receives and validates the rotation-to-stationary traffic: run the GCIFA test, read its fault LEDs and locate the error type in the data-transfer error table. Then run the ROTSOT test to cover the optical chain out to the OPCONTA. If images are also affected, use the Console test, which outputs test data from the GCIFA to tell a gantry-side fault from a console-side one..
Toshiba / Canon Medical CT systems using the MUDAT optical link (Aquilion RX, Aquilion ONE, Aquilion S16, Aquilion 16 and Asteion Super4). ROTSOT designates transmission from the rotation section to the stationary section; the GCIFA runs the error check on that direction and adds extra data to the DAS data before handing it to the preprocessing path.
Full troubleshooting for B710000B
The complete diagnosis, cause tree, and step-by-step remedy for the MUDAT (data transfer system) subsystem are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with real figures.
Open the Activion16 service manualRelated MUDAT (data transfer system) codes
Related service procedures
Source: Toshiba/Canon CT service error message table (errmsg.eng); Aquilion 16 service manual, MUDAT / IF-unit chapter (2D201-059E). Reference information re-authored by FSELIB — not a copy of the OEM manual.