Canon Medical (Toshiba) CT error code, reported by the APPL — positioning status check.

What 90000010 means

On Aquilion RX / V4-62R004 and V3.20R003 and Aquilion ONE / V4.2 CT systems, status code 90000010 indicates: Gantry is tilted — tilt angle not at zero. The code originates in the Gantry (tilt) section of the APPL — positioning status check. Severity level 0 (Information / dialog).

Reported action: Return the gantry tilt to zero before continuing; if the system still does not report the tilt angle as zero, check the gantry tilt mechanism and its angle reporting per the service manual..

Toshiba/Canon Aquilion RX (V4-62R004), V3.20R003 and Aquilion ONE (V4.2) CT systems, on the APPL (application) positioning status check. This code is raised when the gantry is reported tilted; the parallel Japanese-language OEM text for this code states that the gantry tilt angle is not at 0 degrees.

Full troubleshooting for 90000010

The complete diagnosis, cause tree, and step-by-step remedy for the Gantry (tilt) subsystem are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with real figures.

Open the Activion16 service manual

Related Gantry (tilt) codes

Related service procedures

Discuss 90000010 with other field engineers in the FSELIB forum

Source: Toshiba/Canon CT service error message table (errmsg.eng, Aquilion RX V4-62R004 / Aquilion ONE V4.2); cross-referenced with the OEM EMS event-log code table (EMS_Errcode.ini, EMS V3.20.3), including its Japanese-language string, for the same code. Reference information re-authored by FSELIB — not a copy of the OEM manual.