Pular para o conteúdo principal
Information / dialogSHIF

80010006SHIF error — memory pool full

Canon Medical (Toshiba) CT error code, reported by the SHIF.

What 80010006 means

On Aquilion RX / V4-62R004 and V3.20R003 and Aquilion ONE / V4.2 and Aquilion S16 and Asteion Super4 and Aquilion 16 CT systems, status code 80010006 indicates: SHIF error — memory pool full. The code originates in the SHIF section of the SHIF. Severity level 1 (Information / dialog).

Reported action: Retry the operation after the system has settled and note whether the code appears under a specific protocol or after long uninterrupted use. Reboot the console if it persists and call service if the memory pool fills again..

Listed for Aquilion 16, Aquilion RX/V4-62R004, Aquilion ONE/V4.2, Aquilion S16, Asteion Super4 and V3.20R003, under the 30:Device SHIF table, where the OEM text reads "Error in SHIF (Memory pool full)."

Full troubleshooting for 80010006

The complete diagnosis, cause tree, and step-by-step remedy for the SHIF subsystem are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with real figures.

Open the Activion16 service manual

Related SHIF codes

Related service procedures

Discuss 80010006 with other field engineers in the FSELIB forum

Source: errmsg.eng — 30:Device SHIF table. Reference information re-authored by FSELIB — not a copy of the OEM manual.