HC11232 — Pick-up module passed sensor S21 triggered by residual jammed film at startup
Agfa CT error code, reported by the Film Tray Pick-up Module (Front Section).
What HC11232 means
On DryStar 5500 (Type 5364/100) and DryStar 5503 (Type 5364/300) CT systems, status code HC11232 indicates: Pick-up module passed sensor S21 triggered by residual jammed film at startup. The code originates in the Film Tray Pick-up (Front Section) — Passed Sensor S21 section of the Film Tray Pick-up Module (Front Section). Severity level alarm (Alarm).
Reported action: Check the front section and clear the film jam; if it keeps recurring, look for mechanical blockage of the sheet transport and for broken or misaligned sensors..
Raised on the Drystar 5500/5503 pick-up module front section when the pick-up module passed sensor S21 is triggered by a film jammed and still residing in the pick-up module during startup/initialization or the empty-check procedure. Its continuous-activation variant is HC11231.
Full troubleshooting for HC11232
The complete diagnosis, cause tree, and step-by-step remedy for the Film Tray Pick-up (Front Section) — Passed Sensor S21 subsystem are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with real figures.
Open the Activion16 service manualRelated Film Tray Pick-up (Front Section) — Passed Sensor S21 codes
Source: AGFA DryStar service manual — HC error catalogue (FSELIB authored corpus). Reference information re-authored by FSELIB — not a copy of the OEM manual.