HC11231 — Pick-up module passed sensor S21 stuck on (continuous activation) at startup
Agfa CT error code, reported by the Film Tray Pick-up Module (Front Section).
What HC11231 means
On DryStar 5500 (Type 5364/100) and DryStar 5503 (Type 5364/300) CT systems, status code HC11231 indicates: Pick-up module passed sensor S21 stuck on (continuous activation) at startup. The code originates in the Film Tray Pick-up (Front Section) — Passed Sensor S21 section of the Film Tray Pick-up Module (Front Section). Severity level alarm (Alarm).
Reported action: Check the front section and clear the film jam; if it keeps recurring, look for mechanical blockage of the sheet transport and for broken or misaligned sensors..
Raised on the Drystar 5500/5503 pick-up module front section when the pick-up module passed sensor S21 stays continuously activated during startup/initialization or the empty-check procedure. Its dynamic-trigger variant is HC11232.