AlarmFilm Tray Pick-up (Upper) — Passed Sensor S19

HC11229Upper pick-up passed sensor S19 stuck on (continuous activation) at startup

Agfa CT error code, reported by the Film Tray Pick-up Module (Upper Tray).

What HC11229 means

On DryStar 5500 (Type 5364/100) and DryStar 5503 (Type 5364/300) CT systems, status code HC11229 indicates: Upper pick-up passed sensor S19 stuck on (continuous activation) at startup. The code originates in the Film Tray Pick-up (Upper) — Passed Sensor S19 section of the Film Tray Pick-up Module (Upper Tray). Severity level alarm (Alarm).

Reported action: Check the upper input tray and clear the film jam; if it keeps recurring, look for mechanical blockage of the sheet transport and for broken or misaligned sensors..

Raised on the Drystar 5500/5503 upper film tray pick-up module when the upper pick-up passed sensor S19 stays continuously activated during startup/initialization or the empty-check procedure. Its time-out variant is HC11206; the lower-tray equivalent is HC11230.

Full troubleshooting for HC11229

The complete diagnosis, cause tree, and step-by-step remedy for the Film Tray Pick-up (Upper) — Passed Sensor S19 subsystem are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with real figures.

Open the Activion16 service manual

Related Film Tray Pick-up (Upper) — Passed Sensor S19 codes

Discuss HC11229 with other field engineers in the FSELIB forum