HC11206 — Upper pick-up passed sensor S19 not triggered within time-out after pick-up
Agfa CT error code, reported by the Film Tray Pick-up Module (Upper Tray).
What HC11206 means
On DryStar 5500 (Type 5364/100) and DryStar 5503 (Type 5364/300) CT systems, status code HC11206 indicates: Upper pick-up passed sensor S19 not triggered within time-out after pick-up. The code originates in the Film Tray Pick-up (Upper) — Passed Sensor S19 section of the Film Tray Pick-up Module (Upper Tray). Severity level alarm (Alarm).
Reported action: Open the upper input tray and remove any jammed sheet; if this recurs, check for mechanical blockage of the sheet transport. If sensor S19 is badly positioned, reposition it; if it is mechanically broken, power off the printer and replace it (Part S19)..
Raised on the Drystar 5500/5503 upper film tray pick-up module when no sheet triggers the upper pickup passed sensor S19 within the time-out limit after a pick-up. Its counterpart on the lower tray is HC11204; the continuous-activation variant of the same sensor is HC11229.
Full troubleshooting for HC11206
The complete diagnosis, cause tree, and step-by-step remedy for the Film Tray Pick-up (Upper) — Passed Sensor S19 subsystem are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with real figures.
Open the Activion16 service manualRelated Film Tray Pick-up (Upper) — Passed Sensor S19 codes