AlarmTH Module — Power Voltage Rail

HC11105Power voltage delivered to THIB for thermal head not OK

Agfa CT error code, reported by the TH Module — Power Voltage Rail.

What HC11105 means

On DryStar 5500 (Type 5364/100) and DryStar 5503 (Type 5364/300) CT systems, status code HC11105 indicates: Power voltage delivered to THIB for thermal head not OK. The code originates in the TH Module — Power Voltage Rail section of the TH Module — Power Voltage Rail. Severity level alarm (Alarm).

Reported action: The source gives no wiring check for this code — treat it as pointing to one of two swaps: power off and replace the power-supply module, or power off and replace the THIB (Thermal Head Interface Board) if THIB's voltage-measurement functionality is at fault..

Raised on the Drystar 5500/5503 when the power voltage delivered to THIB for the thermal head is not OK, shown on the local display as "Call service". The third of three codes (with HC10305 and HC10829) checking the same power-supply-to-THIB-to-thermal-head voltage rail; its wording and reasons are close to HC10305's, with no documented wiring check.

Full troubleshooting for HC11105