AlarmTH Module — Power Voltage Rail

HC10829Thermal head power voltage out of range at THIB

Agfa CT error code, reported by the TH Module — Power Voltage Rail.

What HC10829 means

On DryStar 5500 (Type 5364/100) and DryStar 5503 (Type 5364/300) CT systems, status code HC10829 indicates: Thermal head power voltage out of range at THIB. The code originates in the TH Module — Power Voltage Rail section of the TH Module — Power Voltage Rail. Severity level alarm (Alarm).

Reported action: Check connections at P6 and P7 on THIB, connector P6 on the power supply, and the wiring from PS1/P6 to THIB/P6 and THIB/P7 with a multimeter — reconnect, fix or replace as needed. If the wiring checks out, replace the THIB board..

Raised on the Drystar 5500/5503 when the thermal head power voltage, measured on the THIB (Thermal Head Interface Board), is out of range, shown on the local display as "Call service". Unlike its two sibling voltage-rail codes (HC10305, HC11105), this is the one the source backs with a documented wiring-check path.

Full troubleshooting for HC10829

The complete diagnosis, cause tree, and step-by-step remedy for the TH Module — Power Voltage Rail subsystem are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with real figures.

Open the Activion16 service manual

Related TH Module — Power Voltage Rail codes

Discuss HC10829 with other field engineers in the FSELIB forum

Source: AGFA DryStar service manual — HC error catalogue (FSELIB authored corpus). Reference information re-authored by FSELIB — not a copy of the OEM manual.