AlarmThermal Head Module — Positioning Mechanics

HC11103Thermal head fails to reach the print position

Agfa CT error code, reported by the Thermal Head Module.

What HC11103 means

On DryStar 5500 (Type 5364/100) and DryStar 5503 (Type 5364/300) CT systems, status code HC11103 indicates: Thermal head fails to reach the print position. The code originates in the Thermal Head Module — Positioning Mechanics section of the Thermal Head Module. Severity level alarm (Alarm).

Reported action: Check the drum-positioning unit for an obstruction and remove it. If the thermal head still fails to reach the print position, power off the printer and replace the TH-positioning motor M3..

Raised on the Drystar 5500/5503 when the thermal head fails to reach the 'print' position. Classed as a Defect, and the printer shows "Call service". Documented in the appendix alongside the counterpart 'home' position fault HC11104.

Full troubleshooting for HC11103