HC11103 — Thermal head fails to reach the print position
Agfa CT error code, reported by the Thermal Head Module.
What HC11103 means
On DryStar 5500 (Type 5364/100) and DryStar 5503 (Type 5364/300) CT systems, status code HC11103 indicates: Thermal head fails to reach the print position. The code originates in the Thermal Head Module — Positioning Mechanics section of the Thermal Head Module. Severity level alarm (Alarm).
Reported action: Check the drum-positioning unit for an obstruction and remove it. If the thermal head still fails to reach the print position, power off the printer and replace the TH-positioning motor M3..
Raised on the Drystar 5500/5503 when the thermal head fails to reach the 'print' position. Classed as a Defect, and the printer shows "Call service". Documented in the appendix alongside the counterpart 'home' position fault HC11104.
Full troubleshooting for HC11103
The complete diagnosis, cause tree, and step-by-step remedy for the Thermal Head Module — Positioning Mechanics subsystem are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with real figures.
Open the Activion16 service manualRelated Thermal Head Module — Positioning Mechanics codes
Source: AGFA DryStar service manual — HC error catalogue (FSELIB authored corpus). Reference information re-authored by FSELIB — not a copy of the OEM manual.